Measurement Request Form (PRIME LAB USE ONLY)
* Required Field
A copy of this request as well as instructions, will be emailed to you and the ICP-OES technician.
Here are the details for the Analyses we currently offer:
- Initial Aluminum Trace Analysis (Al, Be, Fe, Ca, K, Na, Ti, Si) Must be in 5% HF & 1.5% NHO3
- 5 Replicates (measurements) per sample
- High & Low Wavelength Full Frame Images collected for each sample
- Requires a minimum of 14 ml of sample material
- Due to the required sample introduction system for HF samples, results may have an error of upto 15%
- Requires a matrix of 5% HF & 1.5% NHO3
- Final Aluminum Analysis (Al, Be) Must be in 5% NHO3
- 3 Replicates (measurements) per sample
- Requires a minimum of 7 ml of sample material
- Results should have an error of ≤ 1%
- Requires a matrix of 5% NHO3
- Full Suite Analysis
- Requires a minimum of 10 ml of sample material
- Requires a matrix of 5% NHO3
- Full Suite Analysis With Full Frame Images Saved
- High & Low Wavelength Full Frame Images collected for each sample
- Requires a minimum of 14 ml of sample material
- Requires a matrix of 5% NHO3
- PRIME Lab Bio Lab Calcium Interference Analysis (Ca, K, Mg, Na, P, Zn) Must be in 5% NHO3
- 4 Replicates (measurements) per sample
- High & Low Wavelength Full Frame Images collected for each sample
- Requires a minimum of 14 ml of sample material
- Results should have an error of ≤ 1%
- Requires a matrix of 5% NHO3
- Other/Custom
- Custom analyses may be available upon request but not guarenteed
- Submitters may be required to provide their own properly prepaired set of standards
- A series of test samples may be needed to properly fine-tune the instrument setup
- Number of Replicates (measurements) per sample can be set by submitter
- Matrices Can Not contain Organics
- Matrices should NHO3, HF, and or HCl with concentrations ≤ 10%