P_2x5_014_C-1

Production testing

Date:2/12/2007

Operator: Gino Bolla

start testing time:10:19 AM

Sensor: S_2x5_014, Data for leakage current measurement , and plot on overall it is a GOOD sensor
ROCROC waferLabel
ROC0XE4K3KT05B
ROC1XE4K3KT14A
ROC2XE4K3KT13B
ROC3XE4K3KT13A
ROC4XE4K3KT11B
ROC5XE4K3KT18A
ROC6XE4K3KT18B
ROC7XE4K3KT15D
ROC8XE4K3KT17D
ROC9XE4K3KT12A

Depletion Voltage is -50 and suggested operational voltage is -90
Ibias_DAC, Ibias_DAC Data 132 134 146 130 136 150 126 138 126 128
Vana, VanaData 125 135 139 117 141 158 122 141 137 131
CalDelData and plots967387801047390729693
VthrCompData and plots63847475758878807366
TestDdata ,TestD Plots006416000000
LightData ,Light Plots006428000000
Pass/Fail TestD GOOD GOOD BAD GOOD GOOD GOOD GOOD GOOD GOOD GOOD
Pass/Fail LIGHT GOOD GOOD BAD GOOD GOOD GOOD GOOD GOOD GOOD GOOD

Noise MAP

[IMAGE] of

Efficiency MAP

[IMAGE] of

ROC0 NOISE[IMAGE] of [IMAGE] of MEAN = 4.308027

SIGMA = 0.348743
ROC1 NOISE[IMAGE] of [IMAGE] of MEAN = 4.468048

SIGMA = 0.370713
ROC2 NOISE[IMAGE] of [IMAGE] of MEAN = 4.607277

SIGMA = 0.377017
ROC3 NOISE[IMAGE] of [IMAGE] of MEAN = 4.484344

SIGMA = 0.346514
ROC4 NOISE[IMAGE] of [IMAGE] of MEAN = 5.308956

SIGMA = 0.406877
ROC5 NOISE[IMAGE] of [IMAGE] of MEAN = 4.925310

SIGMA = 0.387356
ROC6 NOISE[IMAGE] of [IMAGE] of MEAN = 4.735536

SIGMA = 0.375003
ROC7 NOISE[IMAGE] of [IMAGE] of MEAN = 4.959496

SIGMA = 0.418565
ROC8 NOISE[IMAGE] of [IMAGE] of MEAN = 5.318762

SIGMA = 0.422759
ROC9 NOISE[IMAGE] of [IMAGE] of MEAN = 4.991266

SIGMA = 0.444443

Comment (if any):

This Device have been tested 1Times prior to this one and here are the links to the previous tests:

P_2x5_014_C.html,

End testing time: 10:37 AM