In-plane thermal conductivity measurement
Experimentalist, Michael Wasem
Date of measurement, 7/19/2023
Apparatus, InPlane
Material, INPL22
Sample, 3
Length (mm), 15, 0.05
Width (mm), 50, 0.05
Thickness (mm), 4.11, 0.05
Heater Voltage, 14
Cooler Voltage, 0
Hot Fluxmeter Channels, 105, 106, 107, 108, 109, 110
Cold Fluxmeter Channels, 125, 126, 127, 128, 129, 130
Hot Fluxmeter Temperatures (C), 23.9414, 23.1308, 22.4644, 21.688, 21.0043, 20.1775
Cold Fluxmeter Temperatures (C), 11.2464, 10.4514, 9.7059, 8.98277, 8.25371, 7.40961
Hot Fluxmeter Temperature Uncertainties (C), 0.00120823, 0.00133387, 0.00131687, 0.0013316, 0.00119741, 0.00123216
Cold Fluxmeter Temperature Uncertainties (C), 0.00100596, 0.000854422, 0.00108247, 0.0011003, 0.000888745, 0.000981105
Hot Fluxmeter Temperatures Relative to Hottest Thermistor (C), 0, -0.810047, -1.47699, -2.25323, -2.93717, -3.76404
Cold Fluxmeter Temperatures Relative to Hottest Thermistor (C), 0, -0.794731, -1.54043, -2.26345, -2.99277, -3.83652
Hot Fluxmeter Temperature Relative to Hottest Thermistor Uncertainties (C), 0, 0.00214445, 0.00178439, 0.0017488, 0.00185776, 0.00155823
Cold Fluxmeter Temperature Relative to Hottest Thermistor Uncertainties (C), 0, 0.001722, 0.00108015, 0.00167252, 0.00151476, 0.00105416

Heater Thermistors
Cooler Thermistors
Heater Thermistors relative to hottest thermistor
Cooler Thermistors relative to the hottest thermistor

Analyst, Michael Wasem
Analysis date, 7/20/2023
Apparatus inter-thermistor distance (mm), 8
Apparatus thermistor bore diameter (mm), 1.27
Apparatus fluxmeter width (mm), 50.8
Apparatus fluxmeter thickness (mm), 6.36
Apparatus fluxmeter conductivity (W/mK), 355
Heater Biases: 0, 0, 0, 0, 0, 0
Cooler Biases: 0, 0, 0, 0, 0, 0
Hot fluxmeter flux (W/m^2), 32961 +/- 1251.46
Cold fluxmeter flux (W/m^2), 33615.3 +/- 1276.1
Heat flux through sample (W/m^2), 33288.2 +/- 951.676
Heat flow through sample (W), 10.755 +/- 0.307475
Temperature at hot end of sample (C), 19.4682 +/- 0.189477
Temperature at cold end of sample (C), 11.993 +/- 0.0916879
deltaT (C), 7.47521 +/- 0.210495
R (K/W), 0.695044, 0.0278908
RA (Km^2/W), 0.000142832, 5.99087e-06
Hot flux meter fit
Cold flux meter fit

Hot flux meter difference fit
Cold flux meter difference fit