P_2x3T_041_C-5

Production testing

Date:12/8/2006

Operator: Petra Merkel

start testing time:11:16 AM

Sensor: S_2x3T_04, Data for leakage current measurement , and plot on overall it is a GOOD sensor
ROCROC waferLabel
ROC0XF4K3JT21D
ROC1XF4K3JT24D
ROC2XF4K3JT22C
ROC3XF4K3JT32B
ROC4XF4K3JT32A
ROC5XF4K3JT31B

Depletion Voltage is -10 and suggested operational voltage is -50
Ibias_DAC, Ibias_DAC Data 124 124 126 126 156 133
Vana, VanaData 117 135 127 122 131 125
CalDelData and plots928582907185
VthrCompData and plots796782647572
TestDdata ,TestD Plots00561000