The 129I AMS program at PRIME Lab

Pankaj Sharma, David Elmore, Thomas Miller, Stephan Vogt

Nuclear Instruments and Methods 123 (1997) 347-351

The current status of sample preparation activities and AMS determination of 129I ratios are described. A system background of 20-80 x 10-15 of 129I/I ratio has been achieved without a time-of-flight (TOF) detector and without a low-energy electrostatic deflector. An intercomparison of 129I ratios for AgI samples obtained from other AMS facilities and for the round robin exercise by the International Atomic Energy (IAEA) show an excellent agreement. Potential applications of 129I for tracing groundwater and ocean water are discussed.